TOGO MEDIKIT CO., LTD. SUPERCATH 5 CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Catalog Number SP208-01
Device Problems Fracture (1260); Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 01/17/2022
Event Type  Injury  
Manufacturer Narrative
The device involved in the event was returned. The returned device was examined and the repeatability test (tensile strength testing etc. ) was conducted using reserved samples with the same lot number as that involved in the event. Also, the investigation was conducted by reviewing the records of the manufacturing processes of the iv catheter with the same lot number as that involved in the event, and it was confirmed that there were no manufacturing processes that caused or contributed to the event, and there were no manufacturing records of visual inspections that showed the cause of or contribution to the event. Judging from this examination, a possible cause of this fracture is that repeated bending of the catheter occurred during the indwelling because of insufficient fixation of the device to a patient's body. This resulted in a decrease in tensile strength of the catheter to a point where the catheter could not withstand the pull force and fractured. Lot#: 21k04b1.
 
Event Description
On (b)(6) 2022, at a hospital in (b)(6), it was reported that supercath 5 safety i. V. Catheter was found to be fractured by the confirmation of the device because of the leakage of medical solution from the device during an infusion. There was a possibility that the fractured portion fell off when a fixing tape was peeled off, since the presence of the fractured portion was not confirmed in the patient's body. There was no reported patient injury as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSUPERCATH 5
Type of DeviceCATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
TOGO MEDIKIT CO., LTD.
17148-6 aza kamekawa
oaza hichiya
hyuga city, miyazaki prefecture 883-0 062
JA 883-0062
Manufacturer (Section G)
TOGO MEDIKIT CO., LTD.
17148-6 aza kamekawa
oaza hichiya
hyuga city, miyazaki prefecture 883-0 062
JA 883-0062
Manufacturer Contact
samuel brown
789 n. dixboro road
ann arbor, MI 48105
2024554258
MDR Report Key13581951
MDR Text Key289184014
Report Number9612126-2022-00004
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K140419
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation
Type of Report Initial
Report Date 02/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberSP208-01
Device Lot Number21K04B1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 02/22/2022 Patient Sequence Number: 1
-
-