ZIMMER BIOMET, INC. NEXGEN ARTICULAR SURFACE; PROSTHESIS KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Medical product: kne-unknown--tibial tray, catalog # unknown, lot # unknown, lps-flex fxd mld cd/1-2 14mm, catalog # 00-5964-022-14, lot # 64500611, ng ps mic pre st tib plt sz1 catalog # 00-5980-027-01, lot # 65120836, lps option femoral d-r, catalog # 00-5996-014-52, lot # 64791997, palacos r x 40gms, catalog # 66057851, lot # 60331061.Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the articular surface prosthesis would not seat with mating device.Another device was used to complete the surgery.Attempts to obtain additional information have been made; however, no more is available at this time.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Device was not returned.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d9, g4, g7, h1, h2, h3, h6, h10.H6: health impact - type of investigation - device not returned is n/a which was reported on initial report.Complaint sample was evaluated.Visual evaluation of the returned device shows signs of repeated use and the dovetail feature is flared and compressed.The root cause is attributed to use error.The contact was sent the investigation results.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXGEN ARTICULAR SURFACE
Common Device Name
PROSTHESIS KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13584272
Report Number0001822565-2022-00569
Device Sequence Number20215436
Product Code JWH
UDI-Device Identifier00889024211711
UDI-Public(01)00889024211711(17)270430(10)64377309
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K173057
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00596402212
Device Lot Number64377309
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2022
Supplement Date Received by Manufacturer03/04/2022
09/22/2022
Initial Report FDA Received Date02/23/2022
Supplement Report FDA Received Date03/10/2022
09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/24/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Patient Age56 YR
Patient SexFemale
-
-