|
Model Number CARDIOSAVE HYBRID |
Device Problems
No Display/Image (1183); Fluid Leak (1250); Loss of Power (1475)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 01/21/2022 |
Event Type
malfunction
|
Event Description
|
The rn was just outside the door and heard an unusual sound.
She entered the room and saw the screen had gone black, pump was off and a large puddle was on the floor.
Anesthesia was close by to support the patient with medication until a replacement machine could be set up.
About 450 of the 500 ml bag spilled onto the back of the machine where the ecg and pressure cables connect and then pooled on the floor.
We are guessing that a connection in the transducer tubing came loose and since the heparin was inside a pressure bag at 300mmhg it likely suddenly gushed onto the back of the machine.
Machine was brought to clinical engineering for evaluation.
Manufacturer field service tech opened the machine and found traces of fluid in several spots inside the machine.
It appears fluid reached the power supply board, causing the machine to shut down.
See attached photos.
The ingress protection rating of this machine is specified as ipx0 (no protection against ingress of water).
Power supply board was replaced and unit returned to service.
Manufacturer response for intraorrtic balloon pump, datascope (per site reporter).
Manufacturer field service tech evaluated and repaired the machine.
|
|
Event Description
|
The rn was just outside the door and heard an unusual sound.
She entered the room and saw the screen had gone black, pump was off and a large puddle was on the floor.
Anesthesia was close by to support the patient with medication until a replacement machine could be set up.
About 450 of the 500 ml bag spilled onto the back of the machine where the ecg and pressure cables connect and then pooled on the floor.
We are guessing that a connection in the transducer tubing came loose and since the heparin was inside a pressure bag at 300mmhg it likely suddenly gushed onto the back of the machine.
Machine was brought to clinical engineering for evaluation.
Manufacturer field service tech opened the machine and found traces of fluid in several spots inside the machine.
It appears fluid reached the power supply board, causing the machine to shut down.
The ingress protection rating of this machine is specified as ipx0 (no protection against ingress of water).
Power supply board was replaced and unit returned to service.
Manufacturer response for intraorrtic balloon pump, datascope (per site reporter).
Manufacturer field service tech evaluated and repaired the machine.
|
|
Search Alerts/Recalls
|
|
|