THORATEC CORPORATION HEARTMATE MOBILE POWER UNIT, EU; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number L107758
Device Problems Material Discolored (1170); Loose or Intermittent Connection (1371); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2022
Event Type  malfunction  
Manufacturer Narrative
There was no patient involved in this event.The pma# provided is associated with the device¿s most recent approval.No further information was provided.A supplemental report will be submitted once the manufacturer's investigation is complete.
 
Event Description
It was reported that the mobile power unit (mpu) patient cable was loose at the rubber end around the black and white connectors.The patient cable was also discoloured.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of the rubber casing on the patient cable being loose around the lemo connector and the patient cable on the mobile power unit (mpu) being discolored was confirmed.The mpu was evaluated at the european distribution center (edc) and reported event was confirmed via visual inspection of the returned unit.The mpu was functionally tested and found to function as intended during analysis.The damaged patient cable was replaced.A full functional checkout was performed and the unit passed all tests.The root cause of the reported event could not be conclusively determined through this analysis.Device history records were reviewed and showed no deviations from manufacturing or qa specifications.Heartmate 3 instructions for use section 8 ¿ ¿equipment storage and care¿ and heartmate 3 patient handbook section 6 ¿ ¿caring for equipment¿ explain how to properly care for the equipment, including the mobile power unit (mpu).Heartmate 3 patient handbook section 10 and heartmate 3 instructions for use section f, both entitled ¿safety checklists¿, provide checklists to assist the patient in performing routine maintenance of heartmate 3 lvad, including inspecting the mpu patient cable for damage.This section also informs the user to replace any equipment or system component that appears damaged or worn.Heartmate 3 patient handbook cautions users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE MOBILE POWER UNIT, EU
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key13606259
MDR Text Key289195184
Report Number2916596-2022-01215
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P160054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberL107758
Device Lot Number6916668
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/29/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-