• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Catalog Number SGC0702
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombosis/Thrombus (4440)
Event Date 02/22/2022
Event Type  Injury  
Event Description
This is being filed to report the thrombus requiring aspiration.It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 3-4.After insertion of the steerable guide catheter (sgc) into the left atrium (la), a thrombus was discovered on the guide wire.The wire was pulled back into the dilator and during aspiration both were pulled out of the sgc together.After removing the wire from the dilator, the thrombus could be seen hanging from the wire.The patient was heparinized again and the act was checked more frequently.After this incident there was no thrombus again.The procedure was completed with two clips implanted, reducing mr to 1.There was no clinically significant delay in the procedure and no adverse patient sequela.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Based on the information reviewed, a cause for the reported thrombus cannot be determined; however thrombus is listed in the mitraclip instructions for use (ifu) as a known possible complication associated with mitraclip procedures.The unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13636258
MDR Text Key286352734
Report Number2024168-2022-02084
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/19/2022
Device Catalogue NumberSGC0702
Device Lot Number10916R224
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-