TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
|
Back to Search Results |
|
Model Number 500AHCT |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/10/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during use of the device for cardiopulmonary bypass (cpb), the readings for the partial pressure of carbon dioxide (pco2) values kept dropping.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Event Description
|
Per clinical review: on (b)(6) 2022, the team experienced a problem with the blood parameter monitor (bpm) while on cardiopulmonary bypass whereby the partial pressure carbon dioxide (pco2) values kept dropping despite repeated recalibrations.The unit was changed out, with no delay, no blood loss, and the procedure was completed successfully.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) observed the pco2 value to remain steady throughout the evaluation.No unexpected changes were observed.The unit operated to the manufacturer's specifications.
|
|
Manufacturer Narrative
|
Updated block: h6 the reported complaint could not be confirmed.The service repair technician (srt) was unable to duplicate the reported issue.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|