MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW + ADULT NASAL CANNULA; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW + ADULT NASAL CANNULA; BTT Back to Search Results
Model Number OPT944
Device Problems Stretched (1601); Material Split, Cut or Torn (4008)
Patient Problem Low Oxygen Saturation (2477)
Event Type  malfunction  
Event Description
A healthcare facility in (b)(6) reported via a fisher & paykel healthcare (f&p) field representative that the tubing of an opt944 optiflow + adult nasal cannula had torn whilst on a covid-19 positive patient in the intensive care unit (icu).It was further reported that the patient desaturated to 87% for approximately five minutes and the opt944 device was replaced.No further patient consequences were reported.
 
Manufacturer Narrative
(b)(4).Fisher & paykel healthcare (f&p) are currently in the process of retrieving further information regarding this complaint.Furthermore, the complaint opt944 optiflow + adult nasal cannula is currently en route to f&p for evaluation.We will provide a follow up report upon the completion of our investigation.
 
Manufacturer Narrative
(b)(4).Product background: the opt944 optiflow + adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).Optiflow + interfaces are designed for use with the airvo 2 series of humidification devices.Nhf therapy and the airvo 2 device are intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.The airvo 2 device should not be used for life support purposes, and appropriate patient monitoring must be used at all times.The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Method: the complaint opt944 optiflow + adult nasal cannula was returned to fisher & paykel healthcare (f&p) new zealand for investigation, where it was visually inspected.Results: the customer had stated opt944 optiflow + adult nasal cannula was in use for nine days when it was found torn.The covid-19 patient who was in icu desaturated to 87% spo2 and the cannula was replaced, the patient recovered and there were no further patient consequences.Visual inspection of the returned cannula revealed that the tubing was torn next to the 3-way connector and stretched next to the manifold.Conclusion: we are unable to determine the cause of the reported event.However based on our knowledge of the product and our observation that the tubing was in use for nine days without issue, and was stretched and torn when the incident occurred, the reported event was likely caused by the cannula being subjected to excessive force.Manufacturing controls include inspections during production for visual defects including cracks, tears, inclusions, discoloration and stretching or deformation.Any product that fails the visual inspection is disposed of.The subject opt944 optiflow + adult nasal cannula would therefore have met the required specifications.The user instructions which accompany the opt944 optiflow + adult nasal cannula show in pictorial format the correct placement and fitting of the cannula, including ensuring the headstrap clip and the tubing clip are appropriately attached.The user instructions also warn: "ensure head strap clip is attached, to prevent cannula from being pulled out of the nares." "cannula can become unattached if not used with the head strap clip." "attach tubing clip to clothing/bedding to prevent cannula from pulling off face." "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious.Injury or death." "do not crush or stretch tube, to prevent loss of therapy." "failure to use the set-up described above can compromise performance and affect patient safety.".
 
Event Description
A healthcare facility in ohio reported via a fisher & paykel healthcare (f&p) field representative that the tubing of an opt944 optiflow + adult nasal cannula had torn whilst on a covid-19 positive patient in the intensive care unit (icu).It was further reported that the patient desaturated to 87% for approximately five minutes and the opt944 device was replaced.No further patient consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW + ADULT NASAL CANNULA
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534000
MDR Report Key13654261
MDR Text Key289270143
Report Number9611451-2022-00191
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162553
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT944
Device Catalogue NumberOPT944
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/11/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
F&P PT101 AIRVO 2 HUMIDIFIER.; F&P PT101 AIRVO 2 HUMIDIFIER.
-
-