Model Number IPN001112
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4). There is an investigation ongoing into this reported complaint. A supplement follow-up mdr report will be submitted with the investigation results upon completion of the investigation.
 
Event Description
There was no patient involvement. It was reported that "the machine displayed on the screen an incorrect and fixed value of the helium available in the canister. " as a result, the he regulator assembly was replaced, and the machine was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAC3 OPTIMUS IABP NA/EMEA
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kaylia pen
16 elizabeth drive
chelmsford, MA 01824
MDR Report Key13656415
MDR Text Key286918149
Report Number3010532612-2022-00037
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/01/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/02/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberIPN001112
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/23/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-