MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
|
Back to Search Results |
|
Model Number MMT-751NAB |
Device Problem
Failure to Cycle (1142)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/28/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
Information received by medtronic indicated that the insulin pump had received a compromised force sensor system alarm.Drive support cap was not normal.No harm requiring medical intervention was reported.Troubleshooting was performed; however, the customer will discontinue to use the insulin pump.
|
|
Manufacturer Narrative
|
The device received compromised force sensor system alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform basic occlusion test, occlusion test, prime/compromised force sensor system test, excessive no delivery test due to compromised force sensor system alarm.However, the device passed rewind test and displacement test.The insulin pump history download using thds was successful.There was no compromised force sensor system alarm shown on download report was on event date (b)(6) 2022.No moisture damage found on the electronics, motor, battery tube and vibrator assembly noted.The test p-cap/reservoir does lock into place.The broken battery tube threads, cracked reservoir tube lip, stained address/serial number label and missing end cap sticker were noted during visual inspection.The device received compromised force sensor system alarm during the basic occlusion test due to loose/protruded drive support disk was confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
It was reported that the insulin pump was return for analysis.
|
|
Manufacturer Narrative
|
Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.The information that provided with the initial report was incomplete.The complete information has been included with this report in b5 section.The information related to health effect- clinical code and component code has been updated and provided in h6 section of this report.
|
|