Catalog Number 955468
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/10/2022
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that before use with one unit of prismaflex (b)(4) set c leakage was observed due to a damaged filter port.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The actual device was not available; however, photographs of the sample were provided for evaluation.Visual inspection of the provided photos showed the filter of the set was disconnected from the support plate at the level of the dialysate port.The reported condition was verified.The cause of the event was related to mechanical shock during transportation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX ST100 SET C
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13659891
MDR Text Key286512864
Report Number8010182-2022-00051
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414123468
UDI-Public(01)07332414123468
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Catalogue Number955468
Device Lot Number21E0401
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/21/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-