DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 02/04/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Type of investigation not yet determined: a supplemental report will be submitted upon receipt of additional information.
|
|
Event Description
|
It was reported that during a routine check the cs300 intra-aortic balloon pump (iabp) failed the pneumatic module (pm) leak test.There was no patient involvement, and no adverse event was reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device (10) a getinge field service engineer (fse) was dispatched to evaluate this unit and confirmed the reported malfunction.To address the issue the fse replaced the exch pcb,solenoid driver.The fse confirmed a faulty k5 purge valve and replaced the purge valve assembly,iabp.The unit passed all functional and safety tests to factory specifications.The unit was cleared for clinical use.
|
|
Event Description
|
N/a.
|
|