DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Intermittent Communication Failure (4038)
|
Patient Problems
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/03/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate this unit.The fse discovered that there was intermittent pressure regulator.The fse replaced the pressure regulator and the unit ran without any further incidents.The unit passed all functional and safety tests to factory specifications.The unit was cleared for clinical use.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) had an internal error.There was no harm or injury to the patient and no adverse event was reported.
|
|
Search Alerts/Recalls
|
|
|