TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500 MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
|
Back to Search Results |
|
Model Number 500AHCT |
Device Problem
Incorrect, Inadequate or Imprecise Resultor Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/12/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during use of the device for cardiopulmonary bypass (cpb), the blood parameter monitor (bpm) temperature reading dropped, reading 29-33 degrees celsius.
The surgical procedure was completed successfully.
There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Evaluation is in progress but not yet concluded.
Per the user facility's biomedical engineer, on (b)(6) 2022, the temperature reading on the unit dropped, reading 29-30 degrees celsius.
The unit was flushed and there was no change.
The unit was troubleshot and the arterial cuvette was changed out.
Recalibration of the unit corrected all numbers with the exception of the blood temperature reading.
On (b)(6) 2022, the temperature reading corrected itself and functioned correctly for the next couple of days.
On (b)(6) 2022 the temperature reading dropped again with readings between 29-33 degrees celsius.
The unit was flushed again with no improvement.
Patient was unaffected by the issue.
Both the water heater and the patient temperatures were stable at about 38 degrees celsius and 36-37 degrees celsius respectively.
This complaint is related to (b)(4) / mfr#1828100-2022-00094.
|
|