Model Number MMT-1715KL
Device Problem Mechanical Jam (2983)
Patient Problem Insufficient Information (4580)
Event Date 03/01/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Information received by medtronic indicated that the insulin pump had no motor movement or negligible movement.Retries to free a stuck motor failed alarm.Customer stated that they was able to cleared alarm but unable to complete rewind.No harm requiring medical intervention was reported.Troubleshooting was successfully performed; however, the customer will discontinue use of the device.
 
Manufacturer Narrative
Retainer ring = black.Customer returned insulin pump for an alleged pump error 38 found on (b)(6) 2022.Constant pump error 38 alarms noted during rewind due to corroded motor home switch.Device successfully downloaded to thus.Unable to perform displacement test, prime/seating test, basic occlusion test, force sensor test and occlusion test due to pump error 38 alarms.Device was cut open to perform visual inspection and moisture damage was found on motor.The following were noted during visual inspection: scratched case, pillowing keypad overlay, cracked case (battery tube), corroded motor home switch, and minor scratches on lcd window.In summary, customer allegation for pump error 38 was confirmed due to corroded motor home switch.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715KL 630G
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key13665056
MDR Text Key291740667
Report Number2032227-2022-120701
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000763000316655
UDI-Public(01)000000763000316655(17)230603
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/03/2023
Device Model NumberMMT-1715KL
Device Catalogue NumberMMT-1715KL
Device Lot NumberHG4CT87
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received06/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/03/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
Patient Weight93 KG
-
-