• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH TFNA FENESTRATED HELICAL BLADE 85MM - STERILE ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH TFNA FENESTRATED HELICAL BLADE 85MM - STERILE ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 04.038.385S
Device Problem Migration (4003)
Patient Problem Pain (1994)
Event Date 01/01/2021
Event Type  Injury  
Event Description
Device report from (b)(6) reports an event as follows: it was reported that during a meeting of the aotc held on (b)(6) 2021, slides were presented which represented four tfna cases with medial blade migration into the pelvic area. All four cases were revised to a prosthesis. This captures the case with patient: female, (b)(6) years old at time of event. Patient had begun to complain of pain at the end of (b)(6) 2021. This report is for a tfna fenestrated helical blade 85mm ¿ sterile. This is report 1 of 1 for (b)(4). Additional reports are captured under (b)(4).
 
Manufacturer Narrative
Patient height reported as 160cm. Date of event is an unknown date in 2021. Additional procode: ktt. Complainant part is not expected to be returned for manufacturer review/investigation. The investigation could not be completed; no conclusion could be drawn, as no product was received. Based on the information available, it has been determined that no corrective and/or preventative action is proposed. This complaint will be accounted for and monitored via post market surveillance activities. If additional information is made available, the investigation will be updated as applicable. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameTFNA FENESTRATED HELICAL BLADE 85MM - STERILE
Type of DeviceROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ 4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13666584
MDR Text Key286567298
Report Number8030965-2022-01362
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K160167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number04.038.385S
Device Catalogue Number04.038.385S
Device Lot Number133P361
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/04/2021
Is the Device Single Use? Yes
Type of Device Usage Unkown

Patient Treatment Data
Date Received: 03/04/2022 Patient Sequence Number: 1
Treatment
LOCKSCR 5 L30 F/NAILS TAN LIGHT GREEN; TFNA FEM NAIL Ø10 130° L200 TIMO15
-
-