DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problems
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/08/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse) the cs300 intra-aortic balloon pump (iabp) failed the battery runtime test.There was no patient involvement, and no adverse event was reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device: the fse replaced the batteries to correct the reported issue.Unrelated to the reported issue, the fse replaced the safety disk as it was due to be replaced.Additionally, the fse replaced the solenoid driver board which has been reported under medwatch report 2249723-2022-00449.The fse completed the pm and performed the full calibration, functional testing and electrical safety checks to factory specifications.The unit was returned to the customer and cleared for clinical use.A supplemental report will be submitted upon completion of our investigation.The full name of the initial reporter is (b)(6).The initial reporter named is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(6).
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields; b4, g3, g6, g7, h2, h4, h6, h10, h11 corrected fields: g1.
|
|