• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Premature Discharge of Battery (1057)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2022
Event Type  malfunction  
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse) the cs300 intra-aortic balloon pump (iabp) failed the battery runtime test.There was no patient involvement, and no adverse event was reported.
 
Manufacturer Narrative
Testing of actual/suspected device: the fse replaced the batteries to correct the reported issue.Unrelated to the reported issue, the fse replaced the safety disk as it was due to be replaced.Additionally, the fse replaced the solenoid driver board which has been reported under medwatch report 2249723-2022-00449.The fse completed the pm and performed the full calibration, functional testing and electrical safety checks to factory specifications.The unit was returned to the customer and cleared for clinical use.A supplemental report will be submitted upon completion of our investigation.The full name of the initial reporter is (b)(6).The initial reporter named is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(6).
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields; b4, g3, g6, g7, h2, h4, h6, h10, h11 corrected fields: g1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13673022
MDR Text Key289663691
Report Number2249723-2022-00458
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/13/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-