• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number G34693
Device Problem Peeled/Delaminated (1454)
Patient Problems Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/04/2022
Event Type  malfunction  
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a cook fusion® pre-loaded with acrobat wire guide.It was reported that the hydrophilic coating stripped on the guide wire and as the physician was pulling the wire out, a piece of the hydrophilic coating got caught and broke off in the patient.The plastic stent was placed and then the physician used a cook snare device to retrieve the piece of hydrophilic coating from the patient.The detached piece of hydrophilic coating was retrieved with a cook snare.No further information was provided by the initial reporter, and according to the initial reporter the patient did not require any additional procedures and the patient did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Pma/510(k) # k172288.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.Prior to distribution, all fusion® pre-loaded with acrobat wire guides are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE
Type of Device
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key13685103
MDR Text Key290548217
Report Number1037905-2022-00099
Device Sequence Number1
Product Code KNS
UDI-Device Identifier10827002346939
UDI-Public(01)10827002346939(17)241202(10)W4542552
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG34693
Device Catalogue NumberFS-OMNI-ACRO-35-260
Device Lot NumberW4542552
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
COOK COTTON-HUIBREGTSE BILIARY STENT; COOK SNARE, UNKNOWN MODEL; OLYMPUS TJF-180V ENDOSCOPE
-
-