DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/10/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
At this time, the customer cancelled their request for getinge to evaluate the cs300 intra-aortic balloon pump (iabp) unit involved in this event.A getinge field service engineer (fse) conversed with the customer and was advised that the user switched the rear power switch to the off position.The biomedical engineer switched it to the on position and verified proper charging.Additionally, the unit pumped for 30 minutes.A supplemental report will be submitted if additional information is provided and upon completion of our investigation.
|
|
Event Description
|
It was reported that during routine check, the cs300 intra aortic balloon pump (iabp) unit would not read the ac power, stating that the unit stays on battery.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields; b4, g3, g6, g7, h2, h4, h6, h10, h11 corrected fields: g1.
|
|
Event Description
|
N/a.
|
|