MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problems Mechanical Problem (1384); Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/27/2021
Event Type  malfunction  
Event Description
Information was received from a healthcare provider (hcp) and a company representative regarding a patient who was receiving gablofen (2000 mcg/ml at 979.7 mcg/day) via an implantable pump.The indication for pump use was cerebral palsy and intractable spasticity.On 27-sep-2021 the hcp reported that a safety count alarm occurred.The safety count silent alarm was discussed during the call which resolved the issue.Additional information was received on 29-oct-2021 from the hcp who reported that the logs had been read routinely and then the safety count was noted and discussed with technical support during the initial call.The pump continued to run at the programmed settings and no action was required.Additional information was received on 23-dec-2021 from a company representative who reported that the pump was replaced due to normal battery depletion.
 
Manufacturer Narrative
Analysis of the pump found ¿motor feedthru anomaly ¿ shorting across insulator¿.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13707513
MDR Text Key289017244
Report Number3004209178-2022-03016
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169100824
UDI-Public00643169100824
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial
Report Date 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/14/2016
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/04/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/18/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1579-2013
Patient Sequence Number1
Patient Age30 YR
Patient SexMale
-
-