As reported by user facility: a patient had a code episode and passed away.The machine was removed from service.Pyrogen and bacteriology samples were collected and tested with negative results.The machine was run in simulated treatment mode twice.The dialysate temperature calibration had drifted (34.3°c external meter, machine indicated 37.0°c).The conductivity calibration had also drifted (14.331ms/cm external meter, machine endlf indicated 13.7ms/cm; endlf_s indicated 14.1ms/cm).The ph was 7.5.An electrical safety check was run and no problem was found.There was no level of patient electrical leakage found.
|