• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE 3 SYSTEM CONTROLLER; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE 3 SYSTEM CONTROLLER; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 106531INT
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/10/2019
Event Type  malfunction  
Event Description
It was reported that the white connector of the system controller power cable had a defect.
 
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of damage to the white power cable was not confirmed.The heartmate 3 system controller was not returned for analysis and no log files were submitted for review.Additional information communicated on 01mar2022 indicated that no products would be returning, no information regarding the system controller was available, and no images of the reported damage were available.The root cause of the reported event was not able to be determined in this analysis.Heartmate iii instructions for use section 7 ¿ ¿alarms and troubleshooting¿ and heartmate iii patient handbook section 5 ¿ ¿alarms and troubleshooting¿ explains how to properly interpret and troubleshoot all alarms.Heartmate iii instructions for use section 8 ¿ ¿equipment storage and care¿ and heartmate iii patient handbook section 6 ¿ ¿caring for equipment¿ explain how to properly care for the equipment.The patient handbook cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 SYSTEM CONTROLLER
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key13753256
MDR Text Key287416728
Report Number2916596-2022-01239
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024013297
UDI-Public00813024013297
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P160054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number106531INT
Device Catalogue Number106524US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
Patient Weight81 KG
-
-