MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problems Excess Flow or Over-Infusion (1311); Infusion or Flow Problem (2964); Positioning Problem (3009)
Patient Problem Syncope/Fainting (4411)
Event Date 03/08/2022
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 8780, serial#: (b)(4), implanted: (b)(6) 2012, explanted: (b)(6) 2022, product type: catheter. Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), ubd: 18-aug-2014, udi#: (b)(4). If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider (hcp) via a company representative regarding a patient receiving fentanyl (9000 mcg/ml at 1962 mcg/ml) and clonidine (300 mcg/ml at 81 mcg/day) via an implanted pump. It was reported that the patient was scheduled for a pump replacement. Per the patient, he had multiple incidents of fainting and he believed it was from the pump pumping too much medication intermittently. It was noted that the physician was skeptical of the story. Pump interrogation was as expected. The pump was replaced for normal eri (elective replacement indicator). During the procedure, the hcp could not aspirate through the side port, so the coiled pump segment behind the pump was removed and a new pump segment was added to the remainder of the existing spinal catheter. There were no environmental, external, or patient factors that may have led or contributed to the issue. The issue was noted to be resolved, and it was indicated that the hcp had no further information to provide regarding the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13756659
MDR Text Key287110356
Report Number3004209178-2022-03186
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/28/2019
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received06/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/04/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 03/14/2022 Patient Sequence Number: 1
-
-