• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS M & ST DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS M & ST DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955467
Device Problems Fluid Leak (1250); Material Perforation (2205)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2022
Event Type  malfunction  
Manufacturer Narrative
A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. The device was not available, however, the customer provided photos for evaluation. Visual inspection observed that the access line was perforated resulted in the leak. The reported condition was verified. The cause could not be determined. Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during priming with a prismaflex (b)(4) set, an external fluid leak was observed due to a damaged access tube. There was no patient involvement. No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePRISMAFLEX SETS M & ST
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR 69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13760143
MDR Text Key287185587
Report Number8010182-2022-00064
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 03/14/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number955467
Device Lot Number21F3003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/18/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-