MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problems Retraction Problem (1536); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2022
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 8780, serial#: (b)(4), product type: catheter. Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), ubd: 15-dec-2022, udi#: (b)(4). If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider (hcp) via a company representative regarding a patient receiving an unknown medication via an implanted pump. It was reported that during the surgery and placement of the new pump and catheter, the physician was having difficulty advancing the new catheter and getting it to come out of the needle into the intrathecal space. He pulled the catheter back out of the needle and noticed the catheter was punctured by the stylet. The environmental, external, or patient factor that may have led or contributed to the issue was noted to be the physician attempting to advance the catheter during the new implant procedure. The catheter was not implanted. A new catheter was used to complete the procedure with no further issues. The issue was noted to be resolved, and it was indicated that the hcp had no further information to provide regarding the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13768824
MDR Text Key289199346
Report Number3004209178-2022-03273
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Followup
Report Date 06/20/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date05/28/2022
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-