BOSTON SCIENTIFIC CORPORATION AUTOTOME RX 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
|
Back to Search Results |
|
Model Number M00545180 |
Device Problems
Break (1069); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/21/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The complainant indicated that the device is not available for return; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that an autotome rx 44 was used in the sphincter of oddi during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.The physician inserted the autotome rx 44 into the working channel with intact cutting wire.During the procedure, the physician cut the ampulla.When he assessed if further cut was needed, he noticed that the cutting wire was broken.It was reported that nothing else had occurred to cause this to happen.Additionally, no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.Note: photos of the complaint device outside the patient was provided by the customer and showed the cutting wire was broken and bent.
|
|