Model Number 0998-00-0800-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2022
Event Type  malfunction  
Manufacturer Narrative
The initial reporter named in is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(6).Testing of actual/suspected device : the fse replaced the power management board and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during a repair performed by a getinge field service engineer (fse), when the power management board was replaced, it failed to provide a charge to the batteries of the cardiosave intra-aortic balloon pump (iabp).The iabp unit would power on using ac power.This is a failure upon installation of the power management board.There was no patient involved and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13795073
MDR Text Key287326214
Report Number2249723-2022-00547
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2012
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-