BAXTER HEALTHCARE CORPORATION ULTRAFILTER HECHINGEN; SUBSYSTEM, WATER PURIFICATION Back to Search Results
Catalog Number 114604
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/20/2022
Event Type  malfunction  
Event Description
It was reported that during priming with (b)(4), an external fluid leakage was observed.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H1: the actual device was not available; however, photographs of the sample were provided for evaluation.The visual inspection of the two provided photos showed photo one was wet and a small puddle of water was visible on the floor.The second picture showed dried water stains.Based on past investigations, the most likely failure is a crack in the housing nearby the welding zone.The cause is design related.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFILTER HECHINGEN
Type of Device
SUBSYSTEM, WATER PURIFICATION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
holger - crafoord - strasse 26
hechingen D-723 79
GM   D-72379
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13801752
MDR Text Key292976509
Report Number9611369-2022-00044
Device Sequence Number1
Product Code FIP
UDI-Device Identifier07332414114046
UDI-Public(01)07332414114046
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Catalogue Number114604
Device Lot Number1-1802-H-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/25/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
(B)(4)
-
-