DEPUY MITEK LLC US BARREL TORNADO BURR 5.5MM 5PK; ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE
|
Back to Search Results |
|
Model Number 283889 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported by the affiliate in (b)(4) that during an unknown surgery on an unknown date, it was observed that the barrel tornado burr 5.5mm device was blocked and did not rotate.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
|
Manufacturer Narrative
|
Udi: (b)(4).The lot was unknown.The lot number was unknown; therefore, the expiration date, manufacturing site name, and device manufacture date were unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d4: the lot and expiration date were reported as unknown on the initial report; and have been updated accordingly.Udi: (b)(4).
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: h4: the device manufacture date was reported as unknown on the initial report; and has been updated accordingly.Investigation summary: the complaint device was received for evaluation.Upon reviewing the device, it could be observed that the inner shaft is seized, it was not possible to spin to any direction or separate.No anomalies were found on the black connector.A manufacturing record evaluation was performed for the finished device m2106011 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection, this complaint can be confirmed.A possible root cause for the reported issue can be attributed to the procedural variables, such handling of the device or product interaction during procedure; higher speeds applied on the handpiece can increase the temperature of the device's shaft, lack of adequate irrigation and excessive side loading can contribute to a seized inner shaft, as per ifu: adequate irrigation to the tip of the blade or burr is required to cool the blade and prevent accumulation of excised material in the surgical site.Excessive side loading may result in blade wear and degradation, as well as clogging and/or seizing.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
|