• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105ELPKNA ELI LILY PINK; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105ELPKNA ELI LILY PINK; SYRINGE, PISTON Back to Search Results
Model Number MMT-105ELPKNA
Device Problems Break (1069); Connection Problem (2900)
Patient Problem Insufficient Information (4580)
Event Date 02/27/2022
Event Type  malfunction  
Manufacturer Narrative
Serial number: (b)(4).Software version: 3.8.2, color: pink, battery life remaining: <6 months, per visual inspection: cartridge holder is cracked.Under dose window there is a small amount of orange/red liquid on the corners of the rear shell.No physical damage to cap was noted.Customer reports: cartridge holder is cracked.Unit paired successfully to commercial app.Inpen received with leadscrew fully rewound.The leadscrew was not bent, advanced when dosage knob was pressed dialing a dosage and retracted appropriately.No resistance was observed when dosing without a cartridge installed.The screw advanced every time 30.0u was dialed and dosed until the screw reached max extension.In conclusion: inpen received with broken off piece at the cartridge holder.The cartridge holder did not lock in place.The customer complaint of cartridge holder being cracked was confirmed.
 
Event Description
Information received by medtronic indicated that the cartridge holder of inpen was cracked.Troubleshooting was performed however, the customer will discontinue the use of device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105ELPKNA ELI LILY PINK
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
tricha miles
1223 world trade drive, 100
san diego, CA 92128
7635140379
MDR Report Key13821563
MDR Text Key293359186
Report Number3012822846-2022-00250
Device Sequence Number1
Product Code FMF
UDI-Device Identifier000010862088000337
UDI-Public(01)000010862088000337
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-105ELPKNA
Device Catalogue NumberMMT-105ELPKNA
Device Lot NumberA92TA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2022
Date Manufacturer Received02/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age80 YR
Patient Weight145
-
-