BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS HF; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Catalog Number 107142 |
Device Problem
Air/Gas in Device (4062)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 02/21/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during continuous renal replacement therapy (crrt) with a prismax machine equipped with a thermax blood warmer and a prismaflex hf1400 set, an air in blood alarm was generated approximately 5h 45minutes into treatment, and the prismaflex hf1400 set appeared to have clotting.Extracorporeal blood was not returned to the patient and a blood transfusion was performed.A new treatment was started and air was once again observed in the prismaflex hf1400 set after a few hours of treatment at which time ¿foam¿ and ¿air in blood¿ alarms were generated.The treatment was ended and the extracorporeal blood was manually returned to patient.No additional information was provided.
|
|
Manufacturer Narrative
|
Additional information added to d4, g4, h3, h6 and h10.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection observed the presence of air in the return line.The reported condition was verified.The cause could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
|
|