Catalog Number 107142
Device Problem Air/Gas in Device (4062)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 02/21/2022
Event Type  Injury  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during continuous renal replacement therapy (crrt) with a prismax machine equipped with a thermax blood warmer and a prismaflex hf1400 set, an air in blood alarm was generated approximately 5h 45minutes into treatment, and the prismaflex hf1400 set appeared to have clotting.Extracorporeal blood was not returned to the patient and a blood transfusion was performed.A new treatment was started and air was once again observed in the prismaflex hf1400 set after a few hours of treatment at which time ¿foam¿ and ¿air in blood¿ alarms were generated.The treatment was ended and the extracorporeal blood was manually returned to patient.No additional information was provided.
 
Manufacturer Narrative
Additional information added to d4, g4, h3, h6 and h10.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection observed the presence of air in the return line.The reported condition was verified.The cause could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SETS HF
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13825131
MDR Text Key287556173
Report Number8010182-2022-00068
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042938
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number107142
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/20/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PRISMAX MACHINE; THERMAX BLOOD WARMER DISPOSABLE
Patient Outcome(s) Required Intervention;
-
-