Model Number 25483 |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/1901 |
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
|
Event Description
|
The customer reported the device screen is freezing.No patient impact or consequences were reported.
|
|
Event Description
|
The customer reported the device screen is freezing.No patient impact or consequences were reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative: other, other text: the returned device was evaluated.The unit powered on and off using both battery and ac power.An "audio speaker fault" error message triggers after boot up.The touchscreen is not responsive and the device is not functional when this error is present.During internal inspection, a known good speaker was temporarily installed in customer's device and the error message was not present.The customer's speaker was placed back in the device and the failure was present.The customer's speaker was probed and the speaker measured open.The error message is due to defective speaker.
|
|