This is being filed to report during the procedure, cardiac arrest occurred possibly due to air embolism from possibly the steerable guide catheter.
It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with grade 4.
During preparation, the clip delivery system (cds) (cds0701-xtw, 10910r428) had bubbles in the delivery catheter handle.
Attempts to eliminate the air bubbles let to more bubbles.
Therefore, the cds was not used.
A second cds (cds0701-xtw, 10910r426) was prepped successfully and advanced to the mitral valve.
When positioning the clip, the patient blood pressure dropped, and went into cardiac arrest.
Cardiopulmonary resuscitation (cpr) was performed.
The patient was checked for a pericardial effusion, but there was none seen.
The sgc and cds were flushed, and all valves were cleared.
There were no issues noted with the devices.
Medication and extra oxygen were given to the patient.
The patient was stabilized, and the procedure continued.
The patient¿s blood pressure dropped again and heart rate.
The patient was stabilized again, but it was decided to end the procedure as it was suspected the patient had an air embolism but was not confirmed.
Outside the patient anatomy, the cds was inspected and there appeared to be air bubbles in the cds shaft.
It was then thought that the steerable guide catheter (sgc0701, 10915r207) might have had a leak and not noted.
The patient was stable after the procedure.
No additional information was provided.
|