Intuitive surgical, inc.(isi) received the force bipolar instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the failure of broken grip tips.The instrument was found to have a broken grip at the grip base.A piece approximately 0.18" x 0.28" was found to be broken off the yaw pulley.The broken piece was returned with the instrument.No pieces were missing.The root cause of broken instrument grips tip is typically attributed to mishandling/misuse, such as excess force applied to the instrument jaws.System log review confirmed that the force bipolar instrument was last used on (b)(6) 2022 with one use remaining on system sk3387.A site history review was performed and no other complaints related to this product were identified.No image or procedure video was provided for review.This complaint is considered a reportable event due to the following conclusion: it was alleged that the instrument broke and a fragment fell inside the patient during a da vinci assisted procedure.The fragment was retrieved during the same procedure.At this time it is unknown what caused the breakage to occur.
|