MAUDE Adverse Event Report: STANMORE IMPLANTS WORLDWIDE DISTAL FEMUR MINIMALLY INVASIVE GROWER (MIG) - FEMORAL STEM PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STANMORE IMPLANTS WORLDWIDE DISTAL FEMUR MINIMALLY INVASIVE GROWER (MIG) - FEMORAL STEM PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number UNK_STM
Device Problem Fracture (1260)
Patient Problems Pain (1994); Ambulation Difficulties (2544)
Event Date 03/01/2022
Event Type  Injury  
Manufacturer Narrative
The reported device is similar to a device approved for compassionate use in the united states. It was noted that the device is not available for evaluation. Should additional information become available, it will be provided in a supplemental report upon completion of the investigation. Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies. There have been no other similar events for the lot referenced.
 
Event Description
A patient specific prescription form was received for the patient's right distal femoral mig. Noted on the form: prosthetic fracture of femoral stem. Cement on cement stem 110mm long to match patient's femoral curve/ bow. Extra cortical plate - bilateral with ha agluna. Additional information provided by the rep: "the implant fractured at the stem/collar junction. He had mild pain for 2 weeks and then whilst out for a walk he felt his leg ¿give way¿ resulting in the implant fracture. ".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDISTAL FEMUR MINIMALLY INVASIVE GROWER (MIG) - FEMORAL STEM
Type of DevicePROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STANMORE IMPLANTS WORLDWIDE
210 centennial avenue
centennial park, elstree
borehamwood WD6 3 SJ
UK WD6 3SJ
Manufacturer (Section G)
STANMORE IMPLANTS WORLDWIDE
210 centennial avenue
centennial park, elstree
borehamwood WD6 3 SJ
UK WD6 3SJ
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood WD6 3-SJ
UK   WD6 3SJ
2082386500
MDR Report Key13855856
MDR Text Key287668823
Report Number3004105610-2022-00034
Device Sequence Number1
Product Code KRO
Combination Product (y/n)N
Reporter Country CodeEI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation
Type of Report Initial
Report Date 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Expiration Date07/30/2018
Device Catalogue NumberUNK_STM
Device Lot NumberPIN 21110
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/26/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 03/23/2022 Patient Sequence Number: 1
-
-