• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SEASPINE ORTHOPEDICS CORPORATION MARINER PEDICLE SCREW SYSTEM SET SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SEASPINE ORTHOPEDICS CORPORATION MARINER PEDICLE SCREW SYSTEM SET SCREW Back to Search Results
Model Number 41-1010
Device Problem Premature Separation (4045)
Patient Problem Failure of Implant (1924)
Event Type  malfunction  
Event Description
A patient underwent a 3 level spinal surgery from level l3-l5 on (b)(6) 2021 consisting of seaspine's mariner pedicle screw system. Seaspine was made aware on 25 feb 2022 that the two set screws at the l3 level loosened from the construct in the postoperative period.
 
Manufacturer Narrative
Review of the x-ray provided confirms one set screw has loosened; however, the second reported cannot be visually confirmed. No surgical intervention is planned at this time. Review of labeling: possible adverse events: bending, disassembly or fracture of implant and components loosening of spinal fixation implants may occur due to, latent infection, and/or premature loading, possibly resulting in bone erosion, migration or pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMARINER PEDICLE SCREW SYSTEM
Type of DeviceSET SCREW
Manufacturer (Section D)
SEASPINE ORTHOPEDICS CORPORATION
5770 armada dr.
carlsbad CA 92008
Manufacturer (Section G)
SEASPINE ORTHOPEDICS CORPORATION
5770 armada dr.
carlsbad CA 92008
Manufacturer Contact
audrey mudderman
5770 armada dr.
carlsbad, CA 92008
7602165137
MDR Report Key13857946
MDR Text Key297593362
Report Number3012120772-2022-00009
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10889981155100
UDI-Public10889981155100
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K173882
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/17/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number41-1010
Device Catalogue Number41-1010
Was Device Available for Evaluation? No
Date Manufacturer Received02/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-