• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION SPECTRA WAVEWRITER; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION SPECTRA WAVEWRITER; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-1160
Device Problem Unexpected Therapeutic Results (1631)
Patient Problem Inadequate Pain Relief (2388)
Event Date 08/10/2021
Event Type  Injury  
Event Description
It was reported that the patient was experiencing inadequate stimulation.The patient underwent a revision procedure wherein the old ipg was replaced with a magnetic resonance imaging (mri) capable one.The patient was doing well postoperatively.
 
Manufacturer Narrative
Sc-1160 (sn (b)(6).The returned ipg was analyzed, passed all tests performed, and exhibited normal device characteristics.
 
Event Description
It was reported that the patient was experiencing inadequate stimulation.The patient underwent an ipg replacement procedure and was doing well postoperatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA WAVEWRITER
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key13860578
MDR Text Key287674716
Report Number3006630150-2022-01168
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729951254
UDI-Public08714729951254
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date08/27/2020
Device Model NumberSC-1160
Device Catalogue NumberSC-1160
Device Lot Number341234
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/04/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/30/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age48 YR
Patient SexFemale
-
-