Model Number 16-02-85 |
Device Problems
Device Alarm System (1012); Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/22/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Livanova attempted to request additional information to the customer to clarify the issue but no response has been received up to date.Livanova is currently waiting for customer approval to perform service on site.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error message associated to a patient circuit 1 during maintenance.The exact error code was not communicated to livanova.In addition it was reported that patient circuit 2 was leaking.A reportable malfunction cannot be excluded.There was no patient involvement.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: a livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The patient pump was found to be stuck and it was replaced to solve the reported error.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: a livanova field service representative was dispatched to the facility to investigate the device and could not confirm the reported water leak while the error code associated to one of the patient pump could be reproduced.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: the unit is installed since 2013 and a review of the complaint database did not reveal further reported issue related to this heater cooler.Based on available information and taking into account the review of similar events, the most likely root cause of the reported issue is associated to motor bearing damaged by the corrosion, which consequently does not allow the motor shaft to rotate.Possible causes are water infiltration or water formation due to condensation or high temperatures and high level of humidity in the stationary phase (causes related to environmental conditions).
|
|