Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 03/04/2022 |
Event Type
Injury
|
Event Description
|
It was reported that during drilling of the transverse pin, the tip part broke into the patient and the broken piece remains in the patient body.The patient has not experienced any complications since the initial surgery.No revision surgery is planned.Attempts have been made and no further information has been provided.
|
|
Manufacturer Narrative
|
(b)(4).Foreign country: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.H6: mechanical (g04) - drill.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Surgical notes were not provided.X-ray review by mmi states that the visualized portion shows normal fit and alignment.Bone quality appears normal.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|