Block h6 (device codes): medical device problem code a0401 captures the reportable event of cutting wire broken.Block h10: the returned autotome rx 44 was analyzed, and a visual evaluation noted that the cutting wire was broken and it was not returned, consistent with the findings when the distal tip was observed under magnification.The anchor had evidence of remains of the cutting wire, including the welding points.These findings were also consistent with the findings based on the photos from an x-ray inspection.Additionally, the proximal pierce hole was torn.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can tear the proximal pierced hole.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|