MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 BARD POWERPICC; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 BARD POWERPICC; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Device Problem Restricted Flow rate (1248)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2022
Event Type  malfunction  
Event Description
It was reported via ms&s "caller states they are having issues injecting through the powerpiccs.He states they power inject these picc's during procedures but then the injection rate slows down.The power injector is detecting the psi is elevated close to 300 and then slows the injection rate which can affect quality of the procedure.Caller states the doctors do not think they should use powerpicc' s for this procedure but these patients have poor vascular access." ms&s response "explained that some possibilities could be that fibrin can build up in and around the lumen of the picc catheters.This can lead to an increase in the pressure during a procedure.Discussed keeping the contrast at room temperature to maintain proper viscosity during injections.Discussed catheter kinking and improper flushing as other causes that can contribute to these types of pressure issues.Discussed incorrect flushing sequencing related to the type of needless connectors used which can pull blood or contrast back into the tip of the catheter during this type of procedure.Suggested he discuss these issues with the floor nurse, nursing supervisors techs to ensure these potential issues are addressed.".
 
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD POWERPICC
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13895381
MDR Text Key287894609
Report Number3006260740-2022-00958
Device Sequence Number1
Product Code LJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
UNK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/02/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-