• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200616
Device Problems Leak/Splash (1354); Self-Activation or Keying (1557)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that during a shoulder arthroscopy acromioplasty, after the patient was moved to the operating room and the procedure began, the shaver started to operate by itself without surgeon's activation.The surgeon stopped the device and removed the tip, black oil came out of the device and dripped onto the operating table.The procedure was completed with a non-significant delay using a backup device.No further complications were reported.
 
Manufacturer Narrative
H3, h6: the reported device was received for evaluation.A visual inspection found a discolored cable.A functional evaluation was performed on the returned device and a noisy motor was found.The motor/gearbox could not be removed from the housing for further assessment.Self-activation and leaking were not noted during testing.It was determined the device contributed to the reported event.The complaint of self-activation was not confirmed.Factors that could have contributed to the reported event include a failure of a concomitant device.The complaint of black oil leaking was not confirmed.The noisy motor was confirmed, and the root cause was associated with a mechanical component failure.Factors that could have contributed to the failure include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or a short in the power cord.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failures were documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.No containment or corrective actions are recommended at this time.Correction in h6 (medical device problem code).
 
Event Description
It was reported that during a shoulder arthroscopy acromioplasty, after the patient was moved to the operating room and the procedure began, the shaver started to operate by itself without surgeon's activation.The surgeon stopped the device and removed the tip, black oil came out of the device and dripped onto the operating table.The procedure was completed with a non-significant delay using a backup device.No further complications were reported.As per investigation results, noisy motor was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13896532
MDR Text Key287924025
Report Number1643264-2022-00102
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200616
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received06/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age56 YR
Patient SexFemale
-
-