MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC L3W0750 - DUODERM; DRESSING,WOUND,OCCLUSIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC L3W0750 - DUODERM; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Model Number 187660
Device Problems Delivered as Unsterile Product (1421); Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that there was foreign matter beside the product, and which was inside the dressing package.The product was not used by patient.Photograph depicting the issue was received from the complainant.
 
Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: h6 - investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions h10: investigation summary batch record review: the lot 1f03169 was manufactured on 07/01/2021, bodolay manufacturing line, with a total of (b)(4) market units.Complaint investigator performed a batch record review on 08/21/2022, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct.System application product (sap) material (b)(4) and manufacturing order (b)(4).The batch record review supports that there were no discrepancies related to the issue reported.Returned sample evaluation: photo related to the reported problem was available for evaluation.Investigation conclusion the purpose of this investigation is to identify the potential causes for the complaints reported for the malfunction - foreign matter within product, sterile products (e.G.Metal contamination in dressings, particulates within fluids, gels etc) ¿ and foreign matter (e.G.Hairs, insects) within primary pack (sterile products)¿, in products manufactured on manufacturing lines listed, as well as the corresponding corrective and preventive actions required to prevent issue re occurrence.As part of the preliminary investigation for the nonconformance, the impact was determined as major since the failure is a quality attribute detected by existing detection methods and product effectiveness could have an adverse impact.The risk level based on the individual risk acceptability criteria and risk evaluation above is moderate, meaning that risks are broadly accepted but further risk control measures shall be considered if they are feasible and will reduce the risk further.The complaints as described have been reviewed and does not represent a threat to public safety.In addition, as per work instruction (wi), the severity reported for those cases is high.For these reasons, a corrective/preventative action (capa) plan will be generated.After brainstorming and ishikawa, potential root cause to the failure mode was identified.Corrective and preventive actions will be taken for each of the causes identify for problem solution.Root cause(s): method: dirty carts to transport materials dirty trays solution: create cleaning process of trays and cars used to managed materials in cleaning room during the manufacturing process.Method: mixers with residues from previous mixtures solution: include new cleaning process in the clean schedule.Manpower: inadequate transfer of materials.Solution: retrain warehouse and manufacturing personnel in the correct materials transfer from warehouse to the cleaning room per work instructions (wi).Manpower: inadequate electrocuting lamp maintenance solution: retrain the personnel involved in the correct verification of the electrocuting lamp per work instructions (wi).A corrective/preventative action (capa) plan will be generated to track the implementation of these actions and measure the effectiveness in the product and the process.The investigation associated with related event has been approved and is complete.No additional details have been provided to date.Should additional information become available, a follow up report will be submitted.Fda registration number: reporting site: 1049092 manufacturing site: 9618003.
 
Event Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L3W0750 - DUODERM
Type of Device
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR  91000
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR   91000
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 
3365424679
MDR Report Key13926313
MDR Text Key288045036
Report Number9618003-2022-00280
Device Sequence Number1
Product Code NAD
UDI-Device Identifier00768455174843
UDI-Public00768455174843
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number187660
Device Lot Number1F03169
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received08/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-