ST. JUDE MEDICAL FAST-CATH¿ HEMOSTASIS INTRODUCER, CATH-LOCK¿ LOCKING HUB, 12 CM SHEATH, 7.5 F; INTRODUCER, CATHETER
|
Back to Search Results |
|
Model Number 406705 |
Device Problem
Material Separation (1562)
|
Patient Problems
Foreign Body Embolism (4439); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/24/2022 |
Event Type
malfunction
|
Event Description
|
During an atrial fibrillation ablation procedure, the hemostasis valve detached.The procedure was completed and upon removal of the sheath, it was noted that the hemostasis valve had detached from the tip.Fluoroscopy was performed, which revealed the hemostasis valve remained in the patient.The detached portion was removed using a forceps from the same side of the right inguinal region, but it did not reach, and the guidewire preceded from the contralateral side, and removal was successful with a snare.There were no consequences to the patient.The cause of the detachment was likely due to excessive torsional force applied to the sheath.
|
|
Manufacturer Narrative
|
One 7.5f fast-cath introducer sheath was received for evaluation.Visual inspection revealed the sheath tubing had been torn, stretched and twisted into two sections; both sections were returned.The distal section had been kinked at 0.43¿, 2.16¿ and 2.81¿ proximal to the distal tip; the sheath tubing had been torn 4.11¿ proximal to the distal tip.The proximal section had been creased at 0.45¿ and had been torn at 1.18¿ distal to the distal end of the hemostasis hub; the proximal section remained attached to the hemostasis hub.No other visual anomalies were noted.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the sheath detachment is consistent with damage during use.
|
|
Search Alerts/Recalls
|
|
|