DEPUY INTERNATIONAL LTD - 8010379 ASR ACETABULAR IMPLANT 54; ASR HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 999804754
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Adhesion(s) (1695); Pain (1994); Scar Tissue (2060); Distress (2329); Discomfort (2330); Deformity/ Disfigurement (2360); Metal Related Pathology (4530); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 03/10/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the patient presented with elevated ion count and a somewhat painful hip.This was the surgeons words.Surgeon removed the asr cup and head and replaced with a pinnacle sector cup.Doi: (b)(6) 2007.Dor: (b)(6) 2022.Affected side: right hip affected.
 
Manufacturer Narrative
Product complaint (b)(4).Investigation summary :an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
In addition to what previously alleged, records alleges discomfort, permanent injuries, emotional distress, disability, disfigurement and economic damages.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.Device history lot : device history reviews for asr platform have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.
 
Event Description
After review of medical records, the patient was revised due to failed metal on metal hip arthroplasty.Operative notes indicates extensive scar tissue, mild corrosion of the femoral neck with an intact trunnion.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary = an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.Update 29-aug-2022: update received did not have any additional details to be added in investigation.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot = device history reviews for asr platform have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.H10 additional narrative: e3 initial reporter occupation: lawyer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASR ACETABULAR IMPLANT 54
Type of Device
ASR HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13928903
MDR Text Key288028421
Report Number1818910-2022-05637
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040627
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number999804754
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Treatment
ASR ACETABULAR IMPLANT 54; ASR UNI FEMORAL IMPL SIZE 47; LARGE MOD HEAD ADAPT 12/14 +5; UNKNOWN HIP FEMORAL STEM
Patient Outcome(s) Required Intervention;
Patient Age60 YR
Patient Weight86 KG
-
-