Model Number 04.038.295
Device Problem Migration (4003)
Patient Problem Ulcer (2274)
Event Date 02/28/2022
Event Type  Injury  
Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation. Without a lot number the device history records review could not be completed. Product was not returned. Based on the information available, it has been determined that no corrective and/or preventative action is proposed. This complaint will be accounted for and monitored via post market surveillance activities. If additional information is made available, the investigation will be updated as applicable. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2022, the tfna hardware was removed due to the implant failing. The implant had cut through and had created an ulcer on the patient¿s lateral side. The implant was intact and no fragments were created. The patient was revised with a total hip arthroplasty (tha). This report involves one (1) tfna helical blade 95mm. This is report 1 of 3 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameTFNA FENESTRATED HELICAL BLADE 95MM
Type of DeviceROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
ELMIRA
35 airport road
horseheads NY 14845
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13940047
MDR Text Key288122164
Report Number2939274-2022-01115
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number04.038.295
Device Catalogue Number04.038.395
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unkown

Patient Treatment Data
Date Received: 03/29/2022 Patient Sequence Number: 1
Treatment
3.7MM CANNULATED LOCKING SCREW 38MM; 9/130 DEG TI CANN TFNA 235/LEFT - SILE
-
-