Model Number W4TR02
Device Problem Low impedance (2285)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2022
Event Type  malfunction  
Event Description
It was reported that a low impedance lead warning was triggered. It was determined that the lead warning had been triggered in error and the lead impedance had not been out of range. The cardiac resynchronization therapy pacemaker (crt-p) remains in use. No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Concomitant medical products: 429888 lead, implanted: (b)(6) 2018. If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSERENA QUAD CRT-P MRI SURESCAN
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key13942388
MDR Text Key288178399
Report Number3004209178-2022-03952
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00643169735668
UDI-Public00643169735668
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/14/2019
Device Model NumberW4TR02
Device Catalogue NumberW4TR02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 03/29/2022 Patient Sequence Number: 1
Treatment
S53 LEAD, S45 LEAD
-
-