The reported event was confirmed manufacturing related.1 sample were confirmed to exhibit the reported failure.The device had not met specifications.The product was not used for patient treatment.The product caused the reported failure.Visual evaluation of the returned sample noted one unopened (with original packaging), silicone closed wound suction evacuator.Visual inspection of the sample noted foreign material present inside the bulb evacuator appear excess silicone material.This is out of specification, which states, "product and package (kit) must be free from visible loose or embedded foreign matter".A potential root cause for this failure mode could be ¿good manufacturing practices not followed¿.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labeling could not have prevented the reported failure.The actual/suspected device was inspected.
|