• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number ACU0T0
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/10/2022
Event Type  malfunction  
Manufacturer Narrative
Only the used device was returned loose in a large biohazard bag.The plunger lock and lens stop have been removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was retracted to the fill line.The nozzle tip posterior has stress and an aneurysm that began just after the wound guard.The aneurysm travelled along the posterior of the tip and then enlarged to torn damage.Product history records were reviewed, and the documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The root cause cannot be determined for the reported complaint of "injector failure".Only the used device was returned.The plunger was retracted.The plunger position in relation to the lens during advancement cannot be determined.The observed stress lines and tip aneurysm that has torn may suggest that the lens may not have been in the proper position for advancement.The ifu instructs: after the lens has been advanced to the nozzle line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.After confirming the lens is properly positioned and the haptics are folded properly, proceed with lens implantation.Proceeding with implantation of a misfolded haptic or a lens that appears to be ¿out of position¿ can result in a negative outcome.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that during an intraocular lens (iol) implant procedure, there was an injector failure additional information has been received stated that there was a patient contact.There was no patient harm and patient was not hospitalized.The surgeon prognosis was good.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13950188
MDR Text Key288351745
Report Number1119421-2022-00637
Device Sequence Number1
Product Code KYB
UDI-Device Identifier00380652395342
UDI-Public00380652395342
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberACU0T0
Device Lot Number15229196
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DUOVISC VISCOELASTIC SYSTEM
Patient Age67 YR
Patient SexFemale
-
-