ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR
|
Back to Search Results |
|
Model Number ACU0T0 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/10/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Only the used device was returned loose in a large biohazard bag.The plunger lock and lens stop have been removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was retracted to the fill line.The nozzle tip posterior has stress and an aneurysm that began just after the wound guard.The aneurysm travelled along the posterior of the tip and then enlarged to torn damage.Product history records were reviewed, and the documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The root cause cannot be determined for the reported complaint of "injector failure".Only the used device was returned.The plunger was retracted.The plunger position in relation to the lens during advancement cannot be determined.The observed stress lines and tip aneurysm that has torn may suggest that the lens may not have been in the proper position for advancement.The ifu instructs: after the lens has been advanced to the nozzle line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.After confirming the lens is properly positioned and the haptics are folded properly, proceed with lens implantation.Proceeding with implantation of a misfolded haptic or a lens that appears to be ¿out of position¿ can result in a negative outcome.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A nurse reported that during an intraocular lens (iol) implant procedure, there was an injector failure additional information has been received stated that there was a patient contact.There was no patient harm and patient was not hospitalized.The surgeon prognosis was good.No further information available.
|
|
Search Alerts/Recalls
|
|
|