DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE G PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-52
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2022
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device (10/213): during routine morning checks, it was noticed that the balloon pump was not charging batteries despite being connected to mains power, the fault was confirmed on site.To fix the issue the fse replaced the mains cable reel which was faulty.Performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during a routine check performed by the customer it was observed that the batteries on the cardiosave intra-aortic balloon pump (iabp) were not charging.There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE G PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13950321
MDR Text Key289775764
Report Number2249723-2022-00685
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108438
UDI-Public10607567108438
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-52
Device Catalogue Number0998-00-0800-52
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/04/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-