Model Number 380652-50 |
Device Problem
Failure to Sense (1559)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/28/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a da vinci-assisted procedure the customer had issues with instrument recognition on the system arms.The surgeon converted to laparoscopic prior to calling tech support.Technical support associate (tsa) was unable to view logs, system was disconnected.Customer attempted to connect onsite but was unfamiliar with where or onsite wall port was located.There was no patient harm.Intuitive surgical, inc.(isi) followed up with the customer to confirm that the issue was identified during procedure.Ports were placed.Troubleshooting was completed.Technical support was contacted after converting the procedure.No patient demographic information and no other information available.
|
|
Manufacturer Narrative
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported complaint.The reported complaint was not confirmed based on the field evaluation.The fse was unable to duplicate the reported issue with testing on arms 2 and 3 of the patient cart.The system was tested and verified as ready for use.A review of the site's complaint history does not reveal any related complaints involving this product or this event.No image or procedure video was provided for review.This complaint is considered a reportable event due to the following conclusion: the customer converted after the start of the procedure due to instrument recognition issues on the arm.While there was no harm or injury to the patient, system unavailability after the start of a surgical procedure could likely cause or contribute to an adverse event if it were to recur as it may lead to an injury due to the patient¿s inability to tolerate a conversion.
|
|
Search Alerts/Recalls
|
|
|