Model Number 380652-50
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2022
Event Type  malfunction  
Event Description
It was reported that during a da vinci-assisted procedure the customer had issues with instrument recognition on the system arms.The surgeon converted to laparoscopic prior to calling tech support.Technical support associate (tsa) was unable to view logs, system was disconnected.Customer attempted to connect onsite but was unfamiliar with where or onsite wall port was located.There was no patient harm.Intuitive surgical, inc.(isi) followed up with the customer to confirm that the issue was identified during procedure.Ports were placed.Troubleshooting was completed.Technical support was contacted after converting the procedure.No patient demographic information and no other information available.
 
Manufacturer Narrative
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported complaint.The reported complaint was not confirmed based on the field evaluation.The fse was unable to duplicate the reported issue with testing on arms 2 and 3 of the patient cart.The system was tested and verified as ready for use.A review of the site's complaint history does not reveal any related complaints involving this product or this event.No image or procedure video was provided for review.This complaint is considered a reportable event due to the following conclusion: the customer converted after the start of the procedure due to instrument recognition issues on the arm.While there was no harm or injury to the patient, system unavailability after the start of a surgical procedure could likely cause or contribute to an adverse event if it were to recur as it may lead to an injury due to the patient¿s inability to tolerate a conversion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAVINCI XI
Type of Device
PATIENT SIDE CART, 4-ARM
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key13951387
MDR Text Key297539151
Report Number2955842-2022-10816
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874110720
UDI-Public(01)00886874110720
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial
Report Date 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number380652-50
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-